Anton Paar PMA 4 SC | Pensky-Martens Flash Point Tester with Sample Changer
Previous | 4 / 5 | Next |
Anton Paar Pensky-Martens Flash Point Tester with Sample Changer: PMA 4 SC
- The Pensky-Martens closed-cup tester with sample changer measures the flash point of up to 12 samples automatically.
- PMA 4 SC is suitable for flammability applications on both biodiesel and biodiesel blended fuels along with diesel, heating oil and kerosene. It is also commonly used for other potentially flammable liquids.
Key Features
Benefits at a Glance
- A sample changer carrousel to provide up to 12 samples
- Integrated fire-extinguishing system automatically controlled or manually released (requires N2 or CO2 gas)
- The One-Twist swivel-around multi-function head guarantees the safe
- and easy one-hand connection of all sensors and actuators such as the shutter release, stirrer, igniter and detector contacts
- A multi-detector combines the flash point detector and the temperature probe in
- a solid housing with a perfectly adjusted immersion depth
Convenient Operation
- Real-time measurement indication by FPPNet-software
- PC-software FPPNet is included. It is a control and analysis program that simplifies handling, data storage and program transfer. It provides online display of sample and heater block temperature, gradient (also in an XY-diagram) and test status
- Due to the optimized re-cooling time a rationalization of up to 50 % becomes possible
Customized User Flexibility
- Application-based amount of test inserts (optional)
- Standard brass test inserts (optional)
- Stainless steel or small-volume sample cup (optional)
- Stainless steel sample temperature probe Pt100 (optional)
Technical Specifications
Application range | up to 405 °C (°C/°F selectable) |
Test places | 1 to 12 |
Ignition | Gas and electric (interchangeable) |
Heating | According to method or user-defined |
Stirring | According to method or user-defined |
Cooling | Built-in forced air (fan) |
Flash detection | Thermocouple |
Sample temperature | Pt100 glass probe |
Barometric pressure correction | Automatic correction |
Safety | Overheat protection, automatic shut-off, fire-extinguishing system |
Interfaces | Depending on PC configuration |
Gas supply |
Test flame: 50 mbar of propane or butane Fire extinguisher: 6 bar to 12 bar of nitrogen or carbon dioxide |
Power supply | 115 V/230 V, 50 Hz/60 Hz, 1000 W |
Dimensions | 900 mm x 490 mm x 620 mm (W x D x H) |
Weight | 35 kg |
Standard Methods
- ASTM D93 A+B+C
- ISO 2719 A+B+C
- JIS K 2265-3
- IP 34 A+B
Downloads